Exercise Stress Testing
Recovery Phase
In recovery, the treadmill speed and grade or the ergometer load decreases based
on the protocol configuration.
1.
Press the
Recovery
button to advance to the recovery phase.
The clock begins timing the recovery phase. A maximum 12-lead measurement
is taken (if that is part of the selected protocol).
2.
If necessary, adjust the equipment's speed, grade, or load manually.
You can adjust the speed, grade, or load in the following ways. However, once
you press these keys, the pre-programmed protocol becomes inactive: you must
manually control the speed and grade during the remainder of the exercise
and recovery phases.
• Press
Speed W +
(to increase speed) within 5 seconds of your last workload
change.
• Press
Speed W –
(to decrease speed) within 5 seconds of your last workload
change.
• Press
Grade +
(to increase grade) within 5 seconds of your last workload
change.
• Press
Grade –
(to decrease grade) within 5 seconds of your last workload
change.
The following menu items are available during the recovery phase of the test:
Menu Item
Description
Event
Press to display a list of predefined events.
Edit
Press during the recovery or test end phase to allow you to enter
or edit patient data, reason for test termination, or comments.
Measurements
Reestablish the median complex, set the J point, and select the
three leads used to calculate heart rate.
Leads
Select the leads used for Rhythm Lead 1, 2, and 3, All Leads, Lead
Check, or Lead Placement.
Test End Phase
During the text end phase, the system no longer acquires or stores ECG data or
displays workload, speed, or grade information.
1.
Press and hold the
Test end
button to end the test and start the test end phase.
You MUST hold the
Test end
button for more than one second to activate. This
prevents the test from being stopped by an accidental key press.
2.
Select
Reason for termination
or
Comments
to enter information about this
exercise test.
2046275-016B
MAC
®
5500/MAC
®
5500 HD
73
Содержание MAC 5500
Страница 54: ...Preparing the Patient 54 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 62: ...Entering Patient Information 62 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 80: ...Editing Protocols 80 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 98: ...Deleting ECGs and Orders 98 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 167: ...Maintenance 2046275 016B MAC 5500 MAC 5500 HD 167 ...
Страница 168: ...Maintenance 168 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 178: ...Report Formats 178 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 187: ......