Preparing the Patient
Electrode Placement
This section describes various methods for placing electrodes for both resting and
exercise ECGs.
CAUTION:
DELAYED DIAGNOSIS Improper connection of the leadwires will cause inaccuracies
in the ECG.
Ensure the leadwires are connected properly. Trace each leadwire from its
acquisition module label to its colored connector and then to its electrode to
ensure that it is matched to the correct label leadwire connection location.
Resting ECG Placement
The following methods are applicable for resting ECGs.
Standard 12 Lead Placement
AHA Label
IEC Label
Description
1
V1 Red
C1 Red
Fourth intercostal space at the right sternal border
2
V2 yellow
C2 yellow
Fourth intercostal space at the left sternal border
3
V3 green
C3 green
Midway between location 2 and 4
4
V4 blue
C4 brown
Mid-clavicular line in the fifth intercostal space
5
V5 orange
C5 black
Anterior axillary line on the same horizontal level as 4
6
V6 purple
C6 purple
Mid-axillary line on the same horizontal level as 4 and 5
7
LA black
L yellow
Left deltoid
8
LL red
F green
Above left ankle (Alternate placement, upper leg as close
to torso as possible)
9
RL green
N black
Above right ankle (Alternate placement, upper leg as close
to torso as possible)
10
RA white
R red
Right deltoid
2046275-016B
MAC
®
5500/MAC
®
5500 HD
47
Содержание MAC 5500
Страница 54: ...Preparing the Patient 54 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 62: ...Entering Patient Information 62 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 80: ...Editing Protocols 80 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 98: ...Deleting ECGs and Orders 98 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 167: ...Maintenance 2046275 016B MAC 5500 MAC 5500 HD 167 ...
Страница 168: ...Maintenance 168 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 178: ...Report Formats 178 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 187: ......