Maintenance
Cautions
• Follow the cleaning instructions exactly.
• Wring excess disinfectant from wipe before using.
• Never immerse the device, cables, or leadwires in any liquid, as this may corrode
metal contacts and affect signal quality.
• Do not allow fluid to pool around connection pins. If this happens, blot dry with a
soft, lint-free cloth.
• Never use conductive solutions or solutions that contain chlorides, wax, or wax
compounds to clean the device, cables, or leadwires.
• Never use solutions or products that contain any type of Ammonium Chloride such
as, but not limited to:
• Dimethyl Benzyl Ammonium Chloride
• Quaternary Ammonium Chloride solutions
• Abrasive cleaners or solvents of any kind
• Acetone
• Ketone
• Betadine
• Alcohol-based cleaning agents
• Sodium salts
• Never autoclave or steam clean the device, cables, or leadwires.
• Do not use until thoroughly dry.
Cleaning, Disinfecting, and Storing ECG Cables and
Leadwires
In addition to keeping the MAC system clean and in good repair, it is important to keep
the leadwires clean and disinfected. This section provides instructions for cleaning,
disinfecting, and storing ECG cables and leadwires to extend their life and protect
patients.
NOTE:
This information applies to the Multi-Link cable and leadwires. For systems
with the optional KISS system, see the KISS operator’s manual for cleaning and
disinfecting information.
Cleaning and Disinfecting Leadwires
Proper cleaning and disinfecting prolongs the life of cables and leadwires. Failure to
use the proper cleaning solutions or to follow proper procedures can result in the
following:
• Damage or corrosion
• Diminished signal quality
• Product discoloration
• Metal part corrosion
• Brittle wires and connectors
• Reduced cables and leadwires life
2046275-016B
MAC
®
5500/MAC
®
5500 HD
157
Содержание MAC 5500
Страница 54: ...Preparing the Patient 54 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 62: ...Entering Patient Information 62 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 80: ...Editing Protocols 80 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 98: ...Deleting ECGs and Orders 98 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 167: ...Maintenance 2046275 016B MAC 5500 MAC 5500 HD 167 ...
Страница 168: ...Maintenance 168 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 178: ...Report Formats 178 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 187: ......