System Setup
CT Data Guard — Clinical Trial Setup (cont'd.)
Field
Description
Dose Type
Determines whether the dose type is required on each test. Select
Yes
to require the dose type.
If you require dose type, you can edit the list of dose types. You can
have up to 20 types. Each type can be up to 32 characters long.
Additional Questions
Determines whether the response to additional patient questions
are required on each test. Select
Yes
to require additional questions.
If you select
Yes
, you can define up to five additional questions of
10 characters each. For each question, you must define one of the
following response types:
•
Numbers and letters
Allows the entry of any alphanumeric value. Answers can be
up to 17 characters long.
•
Yes or No
Allows only Yes and No answers.
CT Data Guard — Data Guard Features
Field
Description
Prevent editing of
records
Disables editing ECG files in
File Manager
. Select
Yes
to prevent
ECG editing.
Prevent deleting
of untransmitted
records
Disables the deletion of ECGs that have not been transmitted. Select
Yes
to prevent the deletion of untransmitted ECGs.
Enable record
re-transmit
notification
Enables a notification if users attempt to retransmit records that
have already been transmitted to the MUSE CV system. Select
Yes
to enable the notification.
Critical Values Setup
The optional
Critical Values
function allows you to set up custom notifications for
select ECG events. The notification is displayed after the ECG has been acquired, and
you must acknowledge it before proceeding. They also print on the ECG reports. You
can customize the notification text and select the events that trigger the notifications.
For some events, you can also select the specific values that trigger the notification.
After configuring your critical values, you can save the configuration to an SD card,
which you can use to restore the critical value configuration in the case of a system
failure or to quickly configure other systems.
NOTE:
The
Critical Values
option must be activated before you can configure the
notifications. Refer to
“Option Activation” on page 118
for more information.
136
MAC
®
5500/MAC
®
5500 HD
2046275-016B
Содержание MAC 5500
Страница 54: ...Preparing the Patient 54 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 62: ...Entering Patient Information 62 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 80: ...Editing Protocols 80 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 98: ...Deleting ECGs and Orders 98 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 167: ...Maintenance 2046275 016B MAC 5500 MAC 5500 HD 167 ...
Страница 168: ...Maintenance 168 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 178: ...Report Formats 178 MAC 5500 MAC 5500 HD 2046275 016B ...
Страница 187: ......