Product Overview
4.
Press
Login
.
The
Arrhythmia
screen is displayed.
If the
Barcode Reader
option is enabled, a window opens prompting you to
Scan the Patient barcode
.
5.
If the barcode reader prompt is not displayed, press
Cancel
>
More
>
Main Menu
.
Main Screen Power Up Mode
This procedure describes how to navigate to the
Main Menu
after powering on the
system when
Main Screen
is selected for
Power up mode
in
Basic Setup
.
NOTE:
If you need to perform system setup functions, be sure you log in as a user who is
assigned setup editing privileges.
If the system is configured for
Main Screen Power up mode
and does not have
High
Security Mode
enabled, the
Main Menu
is displayed after powering up the system.
You do not need to press any other keys in order to display the
Main Menu
.
If
High Security Mode
is enabled, when the window opens prompting for a user ID and
password, use the following procedure:
1.
In the
User ID
field, enter your user ID.
2.
To move the cursor to the
Password
field, press
Enter
or the
down arrow
on
the
Trimpad
.
3.
In the
Password
field, enter your password.
4.
Press
Login
.
The
Main Menu
is displayed.
Stress ECG Power Up Mode
This procedure describes how to navigate to the
Main Menu
after powering on the
system when
Stress ECG
is selected for
Power up mode
in
Basic Setup
.
NOTE:
If you need to perform system setup functions, be sure you log in as a user who is
assigned setup editing privileges.
If the system is configured for
Stress ECG Power up mode
and
High Security Mode
is not enabled, the
Stress ECG
screen opens on power up. To go to the
Main Menu
,
press
Cancel
>
More
>
Main Menu
.
If
High Security Mode
is enabled, when the window opens prompting for a user ID and
password, use the following procedure:
1.
In the
User ID
field, enter your user ID.
2.
To move the cursor to the
Password
field, press
Enter
or the
down arrow
on
the
Trimpad
.
3.
In the
Password
field, enter your password.
4.
Press
Login
.
The
Stress ECG
screen is displayed.
42
MAC™ 2000
2053535-002K
Содержание MAC 2000
Страница 30: ...Introduction 30 MAC 2000 2053535 002K ...
Страница 44: ...Product Overview 44 MAC 2000 2053535 002K ...
Страница 56: ...Preparing the Patient 56 MAC 2000 2053535 002K ...
Страница 62: ...Entering Patient Information 62 MAC 2000 2053535 002K ...
Страница 88: ...Arrhythmia Mode Recording 88 MAC 2000 2053535 002K ...
Страница 90: ...RR Analysis 90 MAC 2000 2053535 002K ...
Страница 94: ...RR Analysis 94 MAC 2000 2053535 002K ...
Страница 104: ...Stress Testing 104 MAC 2000 2053535 002K ...
Страница 182: ...Maintenance 182 MAC 2000 2053535 002K ...
Страница 212: ...Technical Specifications 212 MAC 2000 2053535 002K ...
Страница 224: ...224 MAC 2000 2053535 002K ...
Страница 225: ......