Compact Airway Modules, E-CAiOVX, E-CAiOV, E-CAiO, E-COVX, E-COV
9 - 23
2062973-004
a.
Keep the gas sampling line connected to the Sampling line connector.
b.
Leave the other end of the gas sampling line open to room air.
c.
Connect the flowmeter to the Oxygen reference gas inlet with a piece of tubing.
d.
Check that the Reference Flow is within the following range:
31 - 45 ml/min with E-CO, E-COV, E-CAiO and E-CAiOV modules
27 - 40 ml/min with E-COVX and E-CAiOVX modules
7.
Zero Valve Operation
Test the zero valve functionality:
a.
Connect the gas regulator to the calibration gas container.
b.
Connect the end of the gas sampling line to the regulator on the gas container.
Leave the regulator overflow port open to room air.
c.
Select
Monitor Setup
>
Service Calibrations
.
d.
Enter the User Name and the Password and press
Enter
to get into the Calibrations
menu.
e.
Select
Gases
.
f.
Start feeding the specified calibration gas. Wait until the gas values shown in the
gas calibration menu rise approximately to the level indicated in the labeling of the
calibration gas container.
NOTE: The gas values in the Gas Calibrations menu is in percentages (%).
g.
Open the zero valve to room air by selecting
Zero valve Off
(zero position).
h.
Check that the CO2, N2O and anesthesia agent values drop back near 0% and the
O2 reading near 21% (room air).
i.
Stop feeding the calibration gas.
j.
Turn the zero valve back to the normal measurement position by selecting
Zero
valve On
(measurement position).
8.
Gas Calibration
Perform gas calibration according to the instructions in section
.
9.
Agent Identification
NOTE: Perform this test only for E-CAiO, E-CAiOV and E-CAiOVX modules.
Check agent ID unreliability:
a.
Feed the specified calibration gas for at least 30 seconds.
b.
Check that the anesthesia agent is identified as Desflurane and the
IDu
value
(=agent ID unreliability) shown in the
Monitor Setup / Service calibrations / Gases
menu is lower than 75.
If the value is higher, repeat the gas calibration and check the value again.
If the value does not drop below 75, contact GE Healthcare for factory repair.
10. Ambient Pressure
Use a barometer to check the operation of the absolute pressure sensor.
•
Check that the ambient pressure value shown in the Gas Calibrations menu does not
differ more than ± 10 mmHg from the value shown by the barometer.
NOTE: The ambient pressure value in the Gas Calibrations menu is in mmHg.
Содержание E-PSMP
Страница 2: ......
Страница 4: ...Module Frames and Modules 2062973 004 For your notes...
Страница 6: ...Module Frames and Modules 1 ii 2062973 004...
Страница 10: ...Module Frames and Modules 1 4 2062973 004 For your notes...
Страница 11: ...2 Hemodynamic Modules E PRESTN E RESTN E PRETN Rev 00 E PSMP E PSM Rev 01 Technical Manual...
Страница 12: ......
Страница 92: ...Module Frames and Modules 2 78 2062973 004 For your notes...
Страница 94: ...Module Frames and Modules 2 80 2062973 004 For your notes...
Страница 95: ...3 Continuous Cardiac Output Module E PiCCO Technical Manual...
Страница 96: ......
Страница 124: ...Module Frames and Modules 3 26 2062973 004 For your notes...
Страница 125: ...4 Cardiac Output and SvO2 Module E COPSv Rev 01 Cardiac Output Module E COP Rev 01 Technical Manual...
Страница 126: ......
Страница 138: ...Module Frames and Modules 4 10 2062973 004 2 Configuration There is no configuration for the E COP or E COPSv modules...
Страница 159: ...Cardiac Output Modules E COPSv E COP 4 31 2062973 004...
Страница 160: ...Module Frames and Modules 4 32 2062973 004 For your notes...
Страница 162: ...Module Frames and Modules 4 34 2062973 004 For your notes...
Страница 163: ...5 Dual Pressure Module E PP Pressure Temp Module E PT Pressure Module E P Technical Manual...
Страница 164: ......
Страница 198: ...Module Frames and Modules 5 32 2062973 004 For your notes...
Страница 199: ...6 Masimo Module E MASIMO Technical Manual...
Страница 200: ......
Страница 202: ...Module Frames and Modules 6 ii 2062973 004...
Страница 207: ...Masimo Module E MASIMO 6 5 2062973 004 2 Configuration There is no configuration for the E MASIMO module...
Страница 218: ...Module Frames and Modules 6 16 2062973 004 For your notes...
Страница 220: ...Module Frames and Modules 6 18 2062973 004 For your notes...
Страница 221: ...7 Nellcor Compatible Module E NSAT E NSATX Technical Manual...
Страница 222: ......
Страница 224: ...Module Frames and Modules 7 ii 2062973 004...
Страница 236: ...Module Frames and Modules 7 12 2062973 004...
Страница 244: ...Module Frames and Modules 7 20 2062973 004 For your notes...
Страница 246: ...Module Frames and Modules 7 22 2062973 004 For your notes...
Страница 247: ...8 Respiratory Modules E sCAiOVX E sCAiOV E sCAiO E sCOVX E sCOV E sCO Technical Manual...
Страница 248: ......
Страница 284: ...Module Frames and Modules 8 34 2062973 004...
Страница 316: ...Module Frames and Modules 8 66 2062973 004 For your notes...
Страница 319: ...9 Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV E CO Technical Manual...
Страница 320: ......
Страница 336: ...Module Frames and Modules 9 14 2062973 004 For your notes...
Страница 353: ...Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV 9 31 2062973 004...
Страница 379: ...Compact Airway Modules E CAiOVX E CAiOV E CAiO E COVX E COV 9 57 2062973 004...
Страница 380: ...Module Frames and Modules 9 58 2062973 004 For your notes...
Страница 381: ...10 Single width Airway Module E miniC Technical Manual...
Страница 382: ......
Страница 391: ...Single width Airway Module E miniC 10 9 2062973 004 2 Configuration There is no configuration for the E miniC module...
Страница 397: ...Single width Airway Module E miniC 10 15 2062973 004 Complete on the Maintenance check form on page 10 36...
Страница 410: ...Module Frames and Modules 10 28 2062973 004...
Страница 419: ...11 Entropy Module E ENTROPY Rev 01 Technical Manual...
Страница 420: ......
Страница 422: ...Module Frames and Modules 11 ii 2062973 004 7 2 Entropy Module E ENTROPY 01 11 17 Maintenance check form 11 19...
Страница 427: ...Entropy Module E ENTROPY Rev 01 11 5 2062973 004 2 Configuration There is no configuration for the E ENTROPY module...
Страница 442: ...Module Frames and Modules 11 20 2062973 004 For your notes...
Страница 443: ...12 EEG Module E EEG and EEG Headbox N EEG Technical Manual...
Страница 444: ......
Страница 446: ...Module Frames and Modules 12 ii 2062973 004 7 3 EEG Headbox N EEG 12 20 Maintenance check form 12 23...
Страница 452: ...Module Frames and Modules 12 6 2062973 004 2 Configuration There is no configuration for the E EEG module...
Страница 468: ...Module Frames and Modules 12 22 2062973 004 For your notes...
Страница 470: ...Module Frames and Modules 12 24 2062973 004 For your notes...
Страница 471: ...13 BIS Module E BIS Rev 01 Technical Manual...
Страница 472: ......
Страница 480: ...Module Frames and Modules 13 6 2062973 004 2 Configuration There is no configuration for E BIS module...
Страница 494: ...Module Frames and Modules 13 20 2062973 004 For your notes...
Страница 495: ...14 NeuroMuscular Transmission Module E NMT Rev 01 Technical Manual...
Страница 496: ......
Страница 504: ...Module Frames and Modules 14 6 2062973 004 2 Configuration There is no configuration for the E NMT module...
Страница 518: ...Module Frames and Modules 14 20 2062973 004 For your notes...
Страница 520: ...Module Frames and Modules 14 22 2062973 004 For your notes...
Страница 521: ...15 F5 5 Module Frame F7 7 Module Frame Technical Manual...
Страница 522: ......
Страница 532: ...Module Frames and Modules 15 8 2062973 004 2 Configuration There is no configuration for the F5 and F7 module frames...
Страница 558: ...Module Frames and Modules 15 34 2062973 004 For your notes...
Страница 560: ...Module Frames and Modules 15 36 2062973 004 For your notes...
Страница 561: ...16 Patient Data Module v2 Technical Manual...
Страница 562: ......
Страница 620: ...Module Frames and Modules 16 56 2062973 004...
Страница 621: ......