Troubleshooting
165
2062973-002
11.8.5 Recorder issues
11.8.6 Acquisition module problems
Problem: an acquisition module does not work with the patient monitor.
Locate first whether the problem is in the patient monitor or in the acquisition module:
1.
Connect another, similar, known good module to the suspect patient monitor and check if
the module works normally:
−
“Yes” => The suspect module is most likely faulty. Refer to Module Frames and
Modules Technical Manual for troubleshooting instructions.
−
“No” => The problem is most likely in the patient monitor. Continue troubleshooting
the problem according to the related troubleshooting chart below.
OR
Connect the suspect acquisition module to another, similar, known good patient monitor and
check if the module works normally:
−
“Yes” => The problem is most likely in the patient monitor. Continue troubleshooting
the problem according to the related troubleshooting chart below.
−
“No” => The suspect module is most likely faulty. Refer to Module Frames and
Modules Technical Manual for troubleshooting instructions.
Problem
Possible cause
Recommended action
Recorder does not work.
Graph location is not configured
correctly.
Check the configuration:
Monitor Setup >
Printing > Devices > Setup.
Recorder board failure.
Check the status of the +5 V REC and +12
V REC voltages, see section
Hardware statistics.
If one of these above voltages is out of
spec, but the +12 V exists in the DC/DC
board, the recorder board is most likely
faulty. Replace the recorder unit.
Recorder does not work.
Recorder failure.
Replace the recorder unit.
CPU board failure
The serial communication for the recorder
is managed by the EMBC section of the
CPU board.
- Check EMBC Frame logs for possible
If the cause is none of the above, the
problem is most likely in the CPU board.
Replace the CPU board.
Recorder works but nothing
appears on the paper.
Paper installed upsidedown.
Turn the paper roll over.
To test which side is active:
- Place the paper on a hard surface and
draw a line with a fingernail - a dark
line will appear on the active (thermal)
side.
Содержание CARESCAPE Monitor B650
Страница 12: ...CARESCAPE Monitor B650 4 2062973 002 For your notes ...
Страница 16: ...CARESCAPE Monitor B650 8 2062973 002 For your notes ...
Страница 46: ...CARESCAPE Monitor B650 38 2062973 002 For your notes ...
Страница 58: ...CARESCAPE Monitor B650 50 2062973 002 For your notes ...
Страница 64: ...CARESCAPE Monitor B650 56 2062973 002 For your notes ...
Страница 136: ...CARESCAPE Monitor B650 128 2062973 002 For your notes ...
Страница 177: ...Troubleshooting 169 2062973 002 ...
Страница 182: ...CARESCAPE Monitor B650 174 2062973 002 For your notes ...
Страница 218: ...CARESCAPE Monitor B650 210 2062973 002 For your notes ...
Страница 224: ...CARESCAPE Monitor B650 216 2062973 002 For your notes ...
Страница 232: ...CARESCAPE Monitor B650 B 4 4 2062973 002 For your notes ...
Страница 239: ......