CARESCAPE Monitor B650
158
2062973-002
11.8 Problems and solutions
11.8.1
Start-up failures
Problem
Possible causes
Recommended actions
Failure to turn on the patient
monitor, when the following
conditions apply:
- The patient monitor is
connected to AC mains.
- The Mains voltage
indicator is not lit. (see
section
)
- None of the power
management LEDs are
lit.(see section
Unable to turn on the patient
monitor when it is powered
from the AC mains. Monitor
battery is not installed.
Power cord is loose.
Ensure that the power cord is connected properly
to the wall outlet and to the patient monitor.
Blown fuses.
Check the status of the fuses and replace them, if
necessary.
- Use only fuses with correct rating.
- If the fuses are blown repeatedly, investigate
the patient monitor carefully for possible short
circuits.
Power cord is faulty.
Check the power cord for wear and damage and
replace if necessary.
The power outlet does not
meet specified
requirements.
Check the power outlet being used:
- Refer to the patient monitor’s supplemental
information manual for power requirements.
- Check the power outlet being used, see section
The cable between the AC/
DC power supply unit and
the DC/DC board is loose
or faulty.
Check that the cable is intact and properly
connected to the AC/DC power supply unit and the
DC/DC board.
NOTE: You can easily check if the cable is properly
connected to the DC/DC board by detaching the
top cover.
The cable between AC/DC
power supply unit and AC
inlet is loose or faulty.
Check that the cable is intact and properly
connected to the AC/DC power supply unit and to
the AC inlet. Ensure the grounding connection is
good:
- Reconnect the cable, if loose.
- Replace the power supply unit, if the cable is
faulty.
The AC/DC power supply is
faulty.
Replace the AC/DC power supply unit.
The DC/DC board is faulty. Replace the DC/DC board.
Содержание CARESCAPE Monitor B650
Страница 12: ...CARESCAPE Monitor B650 4 2062973 002 For your notes ...
Страница 16: ...CARESCAPE Monitor B650 8 2062973 002 For your notes ...
Страница 46: ...CARESCAPE Monitor B650 38 2062973 002 For your notes ...
Страница 58: ...CARESCAPE Monitor B650 50 2062973 002 For your notes ...
Страница 64: ...CARESCAPE Monitor B650 56 2062973 002 For your notes ...
Страница 136: ...CARESCAPE Monitor B650 128 2062973 002 For your notes ...
Страница 177: ...Troubleshooting 169 2062973 002 ...
Страница 182: ...CARESCAPE Monitor B650 174 2062973 002 For your notes ...
Страница 218: ...CARESCAPE Monitor B650 210 2062973 002 For your notes ...
Страница 224: ...CARESCAPE Monitor B650 216 2062973 002 For your notes ...
Страница 232: ...CARESCAPE Monitor B650 B 4 4 2062973 002 For your notes ...
Страница 239: ......