CARESCAPE Monitor B650
114
2062973-002
NOTE: Make sure that at least one other patient monitor is on the network. The other patient
monitor must be in an admitted state and have an active ECG measurement with a simulator
signal.
1.
Disconnect the network cable from the MC Network connector, if connected.
2.
Check that the wireless network symbol and the signal strength symbol are shown
and that the signal strength is adequate.
3.
Select
Data & Pages – Other Patients
and select a patient from the list.
4.
Select
View
and verify that a window with parameters from another patient displays on
the left side of the screen.
5.
Select
Close View
to close the window.
6.
Reconnect the MC Network cable back to the MC Network connector, if applicable.
8.3.11 IX printers
Perform the following test only if the patient monitor is connected to a printer in the IX Network
and you did not print a test page while you configured the IX printer.
1.
Log in to Webmin.
2.
Select
Configuration
>
Printers > Print Test Page
.
3.
Select the IX Printer to test.
4.
Select
Save
.
5.
Verify that the test page was printed to the selected printer.
6.
Repeat steps 3 to 5 for all connected IX printers.
8.3.12 Insite with EXC
Perform the following test only if the remote service is configured and enabled.
•
Contact your local online support center to find out if they can view the patient monitor.
8.3.13 Test completion
Discharge Patient
or
Reset Case
to discard any changes made to the patient monitor
configuration during checkout.
•
Complete the Appendix
Содержание CARESCAPE Monitor B650
Страница 12: ...CARESCAPE Monitor B650 4 2062973 002 For your notes ...
Страница 16: ...CARESCAPE Monitor B650 8 2062973 002 For your notes ...
Страница 46: ...CARESCAPE Monitor B650 38 2062973 002 For your notes ...
Страница 58: ...CARESCAPE Monitor B650 50 2062973 002 For your notes ...
Страница 64: ...CARESCAPE Monitor B650 56 2062973 002 For your notes ...
Страница 136: ...CARESCAPE Monitor B650 128 2062973 002 For your notes ...
Страница 177: ...Troubleshooting 169 2062973 002 ...
Страница 182: ...CARESCAPE Monitor B650 174 2062973 002 For your notes ...
Страница 218: ...CARESCAPE Monitor B650 210 2062973 002 For your notes ...
Страница 224: ...CARESCAPE Monitor B650 216 2062973 002 For your notes ...
Страница 232: ...CARESCAPE Monitor B650 B 4 4 2062973 002 For your notes ...
Страница 239: ......