CARESCAPE Monitor B650
98
2062973-002
Software activation in progress. Do not disconnect any measurement modules or
other peripheral devices, or shut down the monitor until the software activation is
complete. Activation may take up to 10 minutes. The device will automatically restart
once the software activation is complete.
NOTE: When the software is activated first time after a uDOM replacement, the screen
saver does not appear.
7.
Wait until the software activation completes and the patient monitor restarts
automatically.
8.
Verify that the software activation is successful and the patient monitor runs the
activated software.
Activating host software after next case end / discharge
If you want to activate the new host software with a time delay, that is, after the next case end
/ patient discharge, follow this instruction. You can initiate the software activation process with
a delay during an active patient case.
1.
In the
Host Software
window, verify that the software that you activate is listed as
Inactive
in
Current Software State
.
2.
In
Software License
, if applicable, enter the
Activation Code
for the new software
version.
3.
In
Software Activation
, select
After next case end / discharge.
Содержание CARESCAPE Monitor B650
Страница 12: ...CARESCAPE Monitor B650 4 2062973 002 For your notes ...
Страница 16: ...CARESCAPE Monitor B650 8 2062973 002 For your notes ...
Страница 46: ...CARESCAPE Monitor B650 38 2062973 002 For your notes ...
Страница 58: ...CARESCAPE Monitor B650 50 2062973 002 For your notes ...
Страница 64: ...CARESCAPE Monitor B650 56 2062973 002 For your notes ...
Страница 136: ...CARESCAPE Monitor B650 128 2062973 002 For your notes ...
Страница 177: ...Troubleshooting 169 2062973 002 ...
Страница 182: ...CARESCAPE Monitor B650 174 2062973 002 For your notes ...
Страница 218: ...CARESCAPE Monitor B650 210 2062973 002 For your notes ...
Страница 224: ...CARESCAPE Monitor B650 216 2062973 002 For your notes ...
Страница 232: ...CARESCAPE Monitor B650 B 4 4 2062973 002 For your notes ...
Страница 239: ......