Methods 5
5.6.3
Report points
The
Report points
tab lists the report points in the cycle type definition, ordered
as far as possible in the order they will appear in the cycle. Several injection
commands have a predefined set of report points that are added to the list
when the command is included in the cycle type. You can add your own report
points by filling in the details in the empty row at the bottom of the table. A new
empty row is added whenever you create a report point.
Report points are set at defined times in relation to the start or end of injections
in the cycle. Report points that are set outside the time range for the cycle (i.e. a
significant time before the start of the first injection or after the end of the last
injection) will not be created.
Note:
Do not position report points far away from events, so that they lose their
relevance to the event, or so close to an event so that the report point
window overlaps the event itself.
Enter the report point details as follows:
Name
The report point name must be unique within the cycle type.
Choose a name that reflects the function of the report point.
Sec
Enter the time in seconds between the report point and the event.
Before/
After
Choose whether the report point is to be placed before or after
the event.
Start of/
End of
Choose whether the report point is to be placed relative to the
start or end of the injection.
Inject
Choose the injection to which the report point will be related.
Window
Set the window for the report point calculation. The report point
will be placed at the center of the window, and the reported
response will be an average of the response values within the
window. A window of 5 seconds is adequate for most purposes.
Содержание Biacore T100
Страница 1: ...GE Healthcare Biacore T100 Software Handbook ...
Страница 2: ......
Страница 8: ...8 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 13: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 13 Control Software ...
Страница 14: ...14 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 24: ...2 Control Software general features 2 4 File storage 24 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 30: ...3 Manual run 3 4 Ending a manual run 30 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 76: ...4 Application wizards 4 6 Control experiment wizards 76 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 103: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 103 Evaluation Software ...
Страница 104: ...104 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 138: ...7 Data presentation tools 7 4 Report point table 138 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 207: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 207 Appendices ...
Страница 208: ...208 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 214: ...Appendix A Data import and export A 2 Importing data 214 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 236: ...236 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 237: ......