Biacore T100 Software Handbook BR-1006-48 Edition AE
21
Control Software – general features 2
The options
Show Only Current Curve, Show Curves of Same Type
and
Show
All Curves
control which curves are displayed in the sensorgram window. Curve
types distinguish between unsubtracted and reference-subtracted curves.
Choose the
Event Log
option or click on the
Event Log
button at the right of the
toolbar to display the event log window.
Choose the
Wizard Template
or
Method
options to display the wizard or
method definition for the run. You can edit the definition and save it as a new
wizard template or method. You cannot however change the original definition
that is saved together with the result file.
Notebook
opens a notebook window where details of the run may be recorded.
The notebook is only available during a run or for a completed result file. The run
notebook is saved with the result file and can be viewed in the Evaluation
Software.
For some wizard runs and for test tools, the
Wizard Results
option opens a
window showing the results of the run. All other runs are evaluated in the
Evaluation Software.
Sensorgram Markers
controls display of report point and event markers and
labels in the sensorgram window.
2.3.5
Run menu
The options in the
Run
menu are used to start the different types of runs (see
Chapter 3 and Sections 4.1.2 and 5.9.7).
2.3.6
Tools menu
Options in the
Tools
menu control instrument operations outside the context of
runs.
Prime
flushes the flow system with fresh buffer. There is an option to include
Prime
at the beginning of each wizard- or method-based run. Use the menu
option when you want to flush the system at other times (e.g. before a manual
run).
Shutdown
starts the procedure for shutting down the instrument for long
periods of time (more than 4 days). The procedure displays necessary
instructions on the screen. Details of the shutdown procedure are given in the
Biacore T100 Instrument Handbook.
Standby
puts the instrument in standby mode, which maintains a low buffer or
water flow through the flow system for up to 4 days. Leaving the instrument in
standby mode when not in use is generally recommended. The instrument is
Содержание Biacore T100
Страница 1: ...GE Healthcare Biacore T100 Software Handbook ...
Страница 2: ......
Страница 8: ...8 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 13: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 13 Control Software ...
Страница 14: ...14 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 24: ...2 Control Software general features 2 4 File storage 24 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 30: ...3 Manual run 3 4 Ending a manual run 30 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 76: ...4 Application wizards 4 6 Control experiment wizards 76 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 103: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 103 Evaluation Software ...
Страница 104: ...104 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 138: ...7 Data presentation tools 7 4 Report point table 138 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 207: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 207 Appendices ...
Страница 208: ...208 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 214: ...Appendix A Data import and export A 2 Importing data 214 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 236: ...236 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 237: ......