9 Kinetics and affinity analysis
9.8 Predefined models
186
Biacore T100 Software Handbook BR-1006-48 Edition AE
9.8.5
Kinetics – Two State Reaction
This model describes a 1:1 binding of analyte to immobilized ligand followed by
a conformational change that stabilizes the complex. To keep the model simple,
it is assumed that the conformationally changed complex can only dissociate
through the reverse of the conformational change:
A + B = AB = AB*
Note that conformational changes in ligand or complex do not normally give
a response in Biacore.
A good fit of experimental data to the two-state model
should be taken as an indication that conformational properties should be
investigated using other techniques (e.g. spectroscopy or NMR), rather than
direct evidence that a conformational change is taking place.
Rmax2
Analyte binding capacity of the second ligand
(RU)
Rmax2
Conc
Analyte concentration (M)
Conc
tc
Flow rate-independent component of the
mass transfer constant
tc
Flow
Flow rate (µl/min)
f
kt
Mass transfer constant
tc
×
f
1/3
RI
Bulk refractive index contribution in the
sample
RI
Model parameters
Obtained from
ka1
Association rate constant for analyte binding
(M
-1
s
-1
)
Fitted
kd1
Dissociation rate constant for analyte from
the complex (s
-1
)
Fitted
ka2
Forward rate constant for the conformational
change (s
-1
)
Fitted
kd2
Reverse rate constant for the conformational
change (s
-1
)
Fitted
Rmax
Analyte binding capacity of the surface (RU)
Fitted
Conc
Analyte concentration (M)
Provided as
input
tc
Flow rate-independent component of the
mass transfer constant
Fitted
Содержание Biacore T100
Страница 1: ...GE Healthcare Biacore T100 Software Handbook ...
Страница 2: ......
Страница 8: ...8 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 13: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 13 Control Software ...
Страница 14: ...14 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 24: ...2 Control Software general features 2 4 File storage 24 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 30: ...3 Manual run 3 4 Ending a manual run 30 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 76: ...4 Application wizards 4 6 Control experiment wizards 76 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 103: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 103 Evaluation Software ...
Страница 104: ...104 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 138: ...7 Data presentation tools 7 4 Report point table 138 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 207: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 207 Appendices ...
Страница 208: ...208 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 214: ...Appendix A Data import and export A 2 Importing data 214 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 236: ...236 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 237: ......