8 Concentration analysis
8.3 Evaluating calibration-free measurements
150
Biacore T100 Software Handbook BR-1006-48 Edition AE
The fitting procedure is normally fast. If fitting appears to hang for a particular
sample series, you can click
Accept
to accept the current fitting results for that
series or
Abort current
to abandon the series. Click
Abort remaining
if you want
to abandon all remaining series in the evaluation. The fit status is recorded in the
table of results.
Click
Expand all cycles
in the finished result display to display the details of the
sample series in the cycle list.
The results are reported in terms of
Measured Conc
, which is the value
calculated from the fitting, and
Calc Conc
which is obtained by multiplying the
Measured Conc
by the dilution factor to give the concentration in the original
sample. You can change the units for the reported concentration in the header
for the
Measured Conc
column.
Note:
The diffusion coefficient
D
listed in the evaluation results is the value at
the analysis temperature, calculated automatically in the software from
the value at 20°C provided when the assay is run.
8.3.4
Interpreting the results
Consider the following aspects of calibration-free concentration analysis when
interpreting the results:
•
Samples where the QC ratio of the i is less than about 0.13 should be
discarded (see Section 8.3.1). Results for samples where the initial binding
rate at 5 µl/min is outside the range 0.3–15 RU/s should be treated with
caution.
Содержание Biacore T100
Страница 1: ...GE Healthcare Biacore T100 Software Handbook ...
Страница 2: ......
Страница 8: ...8 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 13: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 13 Control Software ...
Страница 14: ...14 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 24: ...2 Control Software general features 2 4 File storage 24 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 30: ...3 Manual run 3 4 Ending a manual run 30 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 76: ...4 Application wizards 4 6 Control experiment wizards 76 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 103: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 103 Evaluation Software ...
Страница 104: ...104 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 138: ...7 Data presentation tools 7 4 Report point table 138 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 207: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 207 Appendices ...
Страница 208: ...208 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 214: ...Appendix A Data import and export A 2 Importing data 214 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 236: ...236 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 237: ......