Burst suppression ratio (BSR)
BSR is defined as the percentage of time of suppressed (isoelectric, flatline) EEG periods during
the last minute of observation. Emergence of burst suppression pattern may indicate very deep
anesthesia, hypothermia or ischemia.
Typically, during general anesthesia, in the absence of requirements for profound levels of
anesthesia, BSR is 0%. Higher levels of burst suppression indicate very deep hypnosis/
unconsciousness level. Burst suppression generally emerges with Entropy values below 40, but
may not appear even with very low Entropy values.
Entropy range guidelines
NOTE: Individual patients may show different values. These are only guidelines.
Entropy in typical general anesthesia
(1) Awake: RE > SE
(2) Induction: RE = SE
(3) Intubation response: RE increases rapidly
(4) Maintenance: RE = SE
(5) Wake-up: RE increases prior to SE
RE
SE
Description
100
90
Awake
60-40
60-40
Low probability of recall, clinically adequate level for
most surgical operations
<40
<40
Deep anesthesia
0
0
Suppressed EEG
Содержание B20
Страница 2: ......
Страница 16: ...B40 B20 Patient Monitor xii ...
Страница 17: ...1 Introduction ...
Страница 18: ......
Страница 27: ...2 System description ...
Страница 28: ......
Страница 58: ...B40 B20 Patient Monitor 2 30 ...
Страница 59: ...3 Installation ...
Страница 60: ......
Страница 69: ...4 Monitoring basic ...
Страница 70: ......
Страница 79: ...5 Alarms ...
Страница 80: ......
Страница 102: ...B40 B20 Patient Monitor 5 22 ...
Страница 103: ...6 Monitor setup ...
Страница 104: ......
Страница 123: ...7 Trends ...
Страница 124: ......
Страница 140: ...B40 B20 Patient Monitor 7 16 ...
Страница 141: ...8 Print and record ...
Страница 142: ......
Страница 153: ...9 Cleaning and care ...
Страница 154: ......
Страница 163: ...10 Troubleshooting ...
Страница 164: ......
Страница 181: ...11 ECG ...
Страница 182: ......
Страница 205: ...12 Impedance respiration ...
Страница 206: ......
Страница 215: ...13 Pulse oximetry ...
Страница 216: ......
Страница 232: ...B40 B20 Patient Monitor 13 16 ...
Страница 233: ...14 Non invasive blood pressure ...
Страница 234: ......
Страница 248: ...B40 B20 Patient Monitor 14 14 ...
Страница 249: ...15 Invasive blood pressure ...
Страница 250: ......
Страница 261: ...16 Temperature ...
Страница 262: ......
Страница 267: ...17 Airway gas ...
Страница 268: ......
Страница 293: ...18 Entropy ...
Страница 294: ......
Страница 316: ...Abbreviations A 12 ...
Страница 318: ...B40 B20 Patient Monitor B 2 ...
Страница 319: ......