B40/B20 Patient Monitor
15-6
•
Invasive pressures need to be zeroed after reconnecting the pressure transducer or
cable, and whenever the patient’s position is changed. If all channels have not been
zeroed, the message ‘InvBP not zeroed’ appears. However, the alarms advance to yellow
and red levels regardless of zeroing.
Invasive Pressures menu
Zero Pressures
Opens a menu to zero both pressures or one of them.
Ventilation Mode
Respiration causes artifacts in invasive pressures. At the end of expiration the artifact is at its
smallest. Select
Spont
for spontaneous respiration and
Contr
for controlled ventilation.
IBPx Setup menu
Label
Start-up labels are Art and CVP. Other labels are IBP1, IBP2, PA, RAP, RVP, LAP, ICP, ABP, UAC
and UVC. Assigning the appropriate label automatically changes other pressure settings
accordingly.
Scale
Scales are assigned by the monitor when the channel is labeled. Scales can also be individually
adjusted between 10 and 300 mmHg in steps of 10.
Digit Format
With the numeric display format you can choose either the Systolic/Diastolic numbers (
S/D
) or
the Mean pressure value (
Mean
) in large size to the screen. You can also choose all values (
S/D/
M)
to the screen. If the label is ICP, also the
CPP
selection is available (see also page
Response
Use the Trim Knob to change the invasive blood pressure averaging time. The available values
are
Normal
and
B-TO-B
(beat-to-beat). With selection
Normal
, normal averaging is used.
Depending on the label, the values are updated approximately every five seconds. With
selection
B-TO-B
, no averaging is used and the values of the last detected pulse are displayed.
These values can change up to three times per second. This feature is useful when it is
necessary to detect fast pressure changes. NOTE: This setting affects only the displayed values,
not the averaging of invasive pressure trends.
Filter Frequency
Measured signal is filtered to remove noise and artifacts. Use the Trim Knob to adjust the filter
between 4 and 22 Hz.
HR Source
If the ECG signal is affected by too much noise for a reliable heart rate calculation, heart rate
can be calculated mechanically from the pressure (Art) or plethysmographic pulse waveform.
The selected heart rate source is displayed above the numerical display of the heart rate. The
color of the heart rate is the same as that of the source parameter.
Auto
selection prioritizes
IBP
Содержание B20
Страница 2: ......
Страница 16: ...B40 B20 Patient Monitor xii ...
Страница 17: ...1 Introduction ...
Страница 18: ......
Страница 27: ...2 System description ...
Страница 28: ......
Страница 58: ...B40 B20 Patient Monitor 2 30 ...
Страница 59: ...3 Installation ...
Страница 60: ......
Страница 69: ...4 Monitoring basic ...
Страница 70: ......
Страница 79: ...5 Alarms ...
Страница 80: ......
Страница 102: ...B40 B20 Patient Monitor 5 22 ...
Страница 103: ...6 Monitor setup ...
Страница 104: ......
Страница 123: ...7 Trends ...
Страница 124: ......
Страница 140: ...B40 B20 Patient Monitor 7 16 ...
Страница 141: ...8 Print and record ...
Страница 142: ......
Страница 153: ...9 Cleaning and care ...
Страница 154: ......
Страница 163: ...10 Troubleshooting ...
Страница 164: ......
Страница 181: ...11 ECG ...
Страница 182: ......
Страница 205: ...12 Impedance respiration ...
Страница 206: ......
Страница 215: ...13 Pulse oximetry ...
Страница 216: ......
Страница 232: ...B40 B20 Patient Monitor 13 16 ...
Страница 233: ...14 Non invasive blood pressure ...
Страница 234: ......
Страница 248: ...B40 B20 Patient Monitor 14 14 ...
Страница 249: ...15 Invasive blood pressure ...
Страница 250: ......
Страница 261: ...16 Temperature ...
Страница 262: ......
Страница 267: ...17 Airway gas ...
Страница 268: ......
Страница 293: ...18 Entropy ...
Страница 294: ......
Страница 316: ...Abbreviations A 12 ...
Страница 318: ...B40 B20 Patient Monitor B 2 ...
Страница 319: ......