background image

Quickstart Guide 

C2 

Protector 

Controller with Remote 

GDS-49 Sensor Transmitter 

Wiring & Initial Setup Instructions  

1)  Install the C2 

Protector 

controller and GDS-49 remote 

sensor transmitter. As with all sensors, avoid locations 

subject to excessive heat, directed water spray or dust. 

A splash guard is recommended on all externally-

mounted sensors.  

2)  Connect the GDS-49 loop output to the C2 Analog Input 

Board as shown.  

3) 

IMPORTANT: 

If the GDS-49 is to be installed 

intrinsically safe, then an approved isolation barrier must 

be wired between the C2 and GDS-49. See drawing on 

back or contact GDS Corp for more information.  

4) Make sure the selected C2 channel is set to “ON” (see 

the C2 manual, section 2.2.4d).  

5) Calibrate the unit and bump test to verify proper 

operation 

GDS-49  

REMOTE SENSOR TRANSMITTER 

Ground Shield at C2 

Controller (Recommended) 

Important: 

 

The GDS-49 output is non-polarized. Either 

wire may be connected to the C2 +24VDC 

terminal.  

 

If moisture is present, do not install junction 

box with unused inlet facing UP! Water WILL 

leak into junction box! 

TB2

R1

R2

TB1

4-20mA INPUTS

CH2

+24VDC POWER OUT

(For Xmtrs.requiring

24VDC; see manual)

CH1

1

2

3

4

1

2

C2 ANALOG INPUT BOARD 

P/N #10-0221 