Gate Bure Rise & Go 56-316 Скачать руководство пользователя страница 3

3

Standing up using the Bure Rise & Go

2

4

1

3

Because standing up or transporting patients always entails risks we recommend that 
two people are always present to assist the user at such times.

1.  Fit the harness as low as possible 

across the hips and bottom with the 
broad upper part around the hips 
and the narrow lower part around 
the bottom. Attach the strap with the 
Velcro fastener and tighten it using the 
steplessly adjustable safety belt. Note 
that the lower part of the harness at-
tached around the patient’s bottom can 
be opened as required, e.g. for visits to 
the WC. 

2.  Place the user’s legs at a 90˚ angle with 

their feet on the floor or platform. Adjust 
the shin supports to fit the user’s shins; 
remember to avoid pressure on the 

patellae and patellar tendons. Refer to 
page 5 for shin support adjustment.

3.  Thread both attachment belts through 

the walker’s frame and the green 
locking devices. Make sure the belts 
are threaded completely through the 
locking devices. NB! Check especially 
that the belts are undamaged.

 
4.  Connect the power rise harness and 

attachment belts (that are secured to 
the walker) and tighten them using the 
loops; the belts must be taught on both 
sides.

Содержание Bure Rise & Go 56-316

Страница 1: ...1 Bure Rise Go Walker and Stand aid with Power Rise Function Instructions for use Bure Rise Go 56 316 Form 2 248 Rev 8 16 18...

Страница 2: ...ing about in everyday life easier thanks to its smart power rise function The Bure Rise Go can be used by anyone who regularly uses a walker The user can remain active while the power rise function is...

Страница 3: ...ached around the patient s bottom can be opened as required e g for visits to the WC 2 Place the user s legs at a 90 angle with their feet on the floor or platform Adjust the shin supports to fit the...

Страница 4: ...the patient The other carer may stand next to the patient 8 In most cases the patient will not achieve a fully upright standing position encourage the patient to achieve a final upright posture indep...

Страница 5: ...tighten knob Always use the stirrup when helping the patient stand Always rest one foot firmly on the extended stirrup in order to prevent the walker from tipping backwards Raising lowering the frame...

Страница 6: ...hat the battery terminal socket should be removed from the control box if the product will be out of use for an extended period more than three weeks This eliminates the risk of ruining the battery as...

Страница 7: ...s Replace handle grips Fit lock knobs plates as required Fit new cable this must always be connected to the control box IP65 classification Connect replace with new remote control as required Refer to...

Страница 8: ...maged parts immediately must be done by experienced personnel The product is only intended for indoor use Dimensions apply to walkers with 125 mm castors Part No Product Width Max user weight Length H...

Отзывы: