37
2.
On the Gas Check 5000is, use the special screwdriver supplied to undo the captive screw on the
batter compartment cover and remove the cover. This is located at the end of the Gas Check at
position (5).
3.
Remove the battery holder. If required, you can carefully unplug the battery holder so that you
can detach the holder from the Gas Check.
4.
Remove the old batteries (4) from the battery holder. Dispose of the batteries (see section 6.2).
5.
Inspect the inside of the battery compartment and battery holder. If the batteries have leaked,
use a cloth dampened with a weak solution of detergent to clean the battery compartment and
battery holder.
6.
Fit four new batteries in the battery holder. Ensure that you fit the batteries in the correct
orientation. Refit the battery holder to the Gas Check.
7.
.On the Gas Check 5000is, refit the battery compartment cover and use the special screwdriver
to tighten the captive screw and secure the cover in place.
8.
Switch on the Gas Check to ensure that the batteries are correctly fitted. If the display remains
blank when you switch on the Gas Check, refer to section 5.3.
CAUTION
The Gas Check 5000is will blow a factory fitted internal protection fuse if the battery clip is
inadvertently reverse connected. The complete instrument will have to be returned to the
supplier for fuse replacement.
CAUTION
Ensure that the screw is located squarely to the threaded insert before tightening else the
battery compartment cover may not be properly secured due to cross threading. The
instrument will then have to be returned to the supplier for replacement of the Gas Check
case and new serial number.