15 Calibration
Step 1
. Use to enter the calibration menu.
WARNING! Incorrect calibration can result in injury or death to personnel.
Put the service tool over the magnetic sensor location closest to the gas cylinder icon
for one second to access the calibration menu.
15.1 Access the calibration menu
Make sure the menu system is unlocked. See
.
Step 2
. Use to continue or to go back.
Step 3.
Calibration menu.
NOTE: If CANCEL was selected, the menu will auto lock after a ten second delay.
WARNING! Use the calibration plug for each calibration or bump test. Incorrect
calibration can result in injury or death to personnel.
The D-Guard² is factory calibrated before dispatch. You must calibrate the D-Guard²
regularly. Calibration adjusts the response of the D-Guard² to known concentrations of
gas at two points, zero and span.
Zero calibration uses zero-grade air or fresh air, which does not contain any trace of
the detectors intended target gas. You can set the concentration of zero calibration
gas. See
Span calibration uses a known concentration of the intended target gas. You can set
the concentration of span calibration gas. See
.