33
6.3
Workflow: Run a Manual Calibration
The purpose of the manual calibration is to fine-tune the calibration the device.
One or more substance calibrations are modified in the calibration
process. Make sure that an appropriate calibration gas source is
connected to the Testgas In socket at the rear of the housing.
A manual calibration only works within certain margins around the
factory calibration. When it varies too much from the factory
calibration the user is informed, and the results of the manual
calibration process are discarded.
1
Click
CAL
to start a Manual
Calibration Measurement
This will open the Calibration
mode dialog
2
After entering the
Calibra-
tion Mode
it is possible to
Edit/Select
the
Substance
Calibration Definitions
or
directly
Start
the Calibration
measurement
using
the
previous Definitions