![GAS BreathSpec Скачать руководство пользователя страница 61](http://html1.mh-extra.com/html/gas/breathspec/breathspec_user-manual_3638769061.webp)
BreathSpec®
– User Manual
61/242
6.4.3 Flow Controls and Heating Modules
Figure 28: Flow and Temperature Control (Example)
The component parameters are displayed and can be modified in the upper section of
the window and in the left section of the window.
Above the spectra chart the flow rates of the drift and carrier gas controlled by the two
electric pressure controllers (EPC 1 and EPC 2) and the temperatures of the heating
elements of the device are managed. Unused elements are greyed out. The current
flow rates and set-points for EPC 1 (drift gas) and EPC 2 (carrier gas) are displayed.
The maximum value for EPC 1 is 500 ml/min. The maximum value for EPC 2 is 150
ml/min (Depending on the column used). The current temperatures and the set points
of the IMS (T1), the column (T2), the 6-Port valve with sample loop (T3) and the inner
heated transferline (T5) are displayed. The set points can be set using these
controllers. The maximum adjustable temperature value is 80 °C. During the cleaning
process the heating modules can reach temperatures up to 100 °C. These will be
displayed as “> 80 °C”. Heating modules can be turned off by decreasing the respective
value to “off”.
Содержание BreathSpec
Страница 1: ...User Manual BreathSpec G A S Gesellschaft f r analytische Sensorsysteme mbH...
Страница 14: ...BreathSpec User Manual 14 242 Transport box 1 piece Transport palett 60 x 80 cm...
Страница 93: ...BreathSpec User Manual 93 242...
Страница 150: ...BreathSpec User Manual 150 242 3 Start measurement with TD start button...
Страница 166: ...BreathSpec User Manual 166 242 3 Wait until the exporting process is completed The measurement files have been exported...
Страница 177: ...BreathSpec User Manual 177 242 4 Device without top cover...
Страница 184: ...BreathSpec User Manual 184 242 3 Top view of installed sample pump unit 4 Remove the plug 5 Loosen front connector...
Страница 191: ...BreathSpec User Manual 191 242 17 Tighten the three screws of the frontplate...
Страница 215: ...BreathSpec User Manual 215 242 10 4Technical drawing Internal Gasflow...