BreathSpec®
– User Manual
137/242
9.3.3 Workflow: Exchanging the inner syringe body
The sampling tube is capped by a disposable syringe body to allow exchanging of all
parts that are in direct contact to breath. The syringe’s Luer tip protrudes from the
handheld sampler (see images below).
The procedure steps are described as follows:
DANGER
•
Biohazard: wear appropriate gloves at all handling steps
•
Take measures to prevent contamination from one subject/patient
to other
WARNING
•
Pinching possible
1
- move click located at the
bottom of the handheld to
disconnect handheld and
tube and pull out the tube.