BreathSpec®
– User Manual
117/242
10
Switch on the
BreathSpec.
7.6 Prepare the device for operation
Before using the device for the first time or after being disconnected from the
nitrogen source for some time it must be cleaned to ensure proper operating
conditions. In this case start the
cleaning mode
.
An appropriate cleaning period duration must be chosen depending on how long the
device was switched off and on the extent of contamination. It is recommended to
clean the device for at least
15 hours
before operating it for the first time. If the
Spectrum is not clean repeat the process. A
reference spectrum
for the evaluation
can be found in the supplied
Analytical Approval
.
INFORMATION!
After the cleaning process the device needs at least
2 hours
to cool
down the internal components to their required temperatures.
The duration of the cool down process is depending on the
temperature default settings.
Содержание BreathSpec
Страница 1: ...User Manual BreathSpec G A S Gesellschaft f r analytische Sensorsysteme mbH...
Страница 14: ...BreathSpec User Manual 14 242 Transport box 1 piece Transport palett 60 x 80 cm...
Страница 93: ...BreathSpec User Manual 93 242...
Страница 150: ...BreathSpec User Manual 150 242 3 Start measurement with TD start button...
Страница 166: ...BreathSpec User Manual 166 242 3 Wait until the exporting process is completed The measurement files have been exported...
Страница 177: ...BreathSpec User Manual 177 242 4 Device without top cover...
Страница 184: ...BreathSpec User Manual 184 242 3 Top view of installed sample pump unit 4 Remove the plug 5 Loosen front connector...
Страница 191: ...BreathSpec User Manual 191 242 17 Tighten the three screws of the frontplate...
Страница 215: ...BreathSpec User Manual 215 242 10 4Technical drawing Internal Gasflow...