GASCOSEEKER 2-500 USER HANDBOOK
64
BACKGROUND ILLUMINATION DOES NOT
OPERATE
IS INFRA RED LINK CLEAN ON INSTRUMENT TOP
PLATE?
The infrared sensor also detects the ambient light condition
and requires the apertures to be clear.
ACTION:
USE A DRY CLOTH TO REMOVE ANY
FOREIGN MATERIAL OR DEBRIS FROM
THE INFRARED APERTURE ON THE
TOP PLATE.
IS BACKGROUND LIGHT LEVEL SUFFICIENTLY LOW?
The background illumination will only be initiated when the
ambient lighting is sufficiently low that the scale cannot be
read.
ACTION:
ENSURE THAT AMBIENT LIGHT IS LOW
ENOUGH IN AREA OF INFRARED
SENSOR.
IS INSTRUMENT BEING USED ON A DEPOT CHECKER,
CONNECTED TO AN ALARM UNIT OR BEING
CALIBRATED?
When instrument is on a DEPOT CHECKER, or is
connected to any other equipment via the infra red link,
then background illumination Is inhibited.
ACTION:
REMOVE INSTRUMENT FROM
CHECKER OR ANY OTHER
CONNECTION.
SWITCH INSTRUMENT OFF.
SWITCH ON AGAIN AND CHECK
ILLUMINATION.