Instructions for Use
Intra-oral Adjustment/Polishing for Zirconia and Lithium Disilicate
Suitability for use
Visually check the polishers for cracks, chips and corrosion.
Ensure that the polishing medium remains secure on the
bur. Do NOT use if damaged or worn out.
Sterilization
Instruments are shipped non-sterile. Complete
cleaning and sterilization instructions may be found at
www.garrisondental.com/sterilization
. Autoclave the
instruments at 134°C with a holding time of fi ve (5)
minutes and a drying time of ten (10) minutes. It is the
responsibility of the dentist or licensed practitioner to
ensure effective sterilization.
Storage
Store the polishers in the sterilization
container until required for use.
Storage should be in dry, clean
conditions at ambient temperature.