Failure
Possible Cause
The display does not
light up.
The balance is not switched on.
The mains supply connection has been interrupted (mains cable not plugged
in/faulty).
Power supply is interrupted.
(Rechargeable) batteries are inserted incorrectly or empty.
No (rechargeable) batteries are inserted.
The displayed weight is
constantly changing
Draught /air movement.
Table / floor vibrations.
Weighing plate has contact with other objects.
Electromagnetic fields / static charging (choose different location/switch off
interfering device if possible).
The weighing result is
obviously wrong
The display of the balance is not at zero.
Adjustment is no longer correct.
Great temperature fluctuations.
Warm-up time was ignored.
Electromagnetic fields / static charging (choose different location/switch off
interfering device if possible).
F----L
Lower zero position of sensor signal (re- Adjustment).
F----H
Weighing capacity is exceeded.
C----L
Lower zero position of sensor signal (re- Adjustment).
C----F
Calibration error, instability or not at zero point during calibrating.
C----H
Calibration error, items on the pan or at higher zero point during calibrating.
chap. “Displayinformation / Error messages”
Should other error messages occur, switch balance off and then on again. If the error message remains
inform manufacturer.