11.
Rear panel
Power input is a standard IEC interface.
Figure 21. QPC rear panel
11.1.
Connector pair, HV output connectors and safeconn connectors
Connector pair J501 - 504/J401 - 404, HV output connectors and
safeconn (HV interlock) connectors
The SHV 10 kV (Fischer) type connector and corresponding SMB (mini
banana) connector is supplied for each high voltage output connection.
The SHV 10 kV (Fischer) type connectors are used to relay high voltage to the
ion pump.
The SMB (mini banana) connections are used to relay safeconn (HV interlock)
logic signal from controller to the ion pump high voltage feedthrough.
When connection HV cable to the controller, insert mating connector into
desired J50X connection and lock the connector.
SMB (mini banana) connectors are push-on connectors. Press firmly until the
connector snaps into place.
Note:
Other connector types available per request.
11.2.
Connector J115, NET (Profibus, optional)
Standard Sub-D 9-pin female connector. Connection to Profibus interface.
For Profibus information reference Profibus user manual, document 900033,
available on Gamma Vacuum website under download section.
11.3.
Connector J113, CTRL
Standard Sub-D 15-pin female connector.
Used in combination with Remote TSP and NEG Controller to facilitate
communications between controllers. Connection is made using straight-
through Sub-D 15-pin cable available from Gamma Vacuum (PN 380049 -
Page 34
900031_J
900031_J - Rear panel