1
2
3
4
5
6
7
8
Set Point
Set Point
Supply 1
Supply 2
Display
Select
Display
Select
5.
As you press each key to test it (press the
ESC
key last), the key is displayed on the
bottom display. Some keys are displayed as they are. Other keys (the
ENTER
key
for example) are displayed as symbols. The keys are displayed as follows:
Keys
Displayed As
1, 2, 3, 4, 5, 6, 7, 8, 9, 0
1, 2, 3, 4, 5, 6, 7, 8, 9, 0
Backspace (
←
) key
ENTER
Menu (blank) key
M
Display Select
1 key
A
Display Select
2 key
B
6.
Press
ESC
to exit the keypad test and enter the LED test. The MPC first tests the
ODD LEDs.
Set Point
Set Point
1
2
3
4
5
6
7
8
Leds flash as they are tested.
7.
Press
ENTER
again test ALL LEDs.
8.
Press
ENTER
again test EVEN LEDs.
9.
Press
ESC
to return to the DISPLY/KEY submenu. Press
ESC
again to step up to
DIAGNOSTICS.
CALIBRATION
For the calibration procedure remove the HV cables from the rear panel of the DIGITEL MPC.
1.
From DIAGNOSTCS, press
ENTER
to step to its submenu. Press
Menu
to cycle to
CALIBRATION.
2.
Press
ENTER
to start calibrating the MPC. A momentary message "TURNING
SUPLYS OFF" appears to let you know that the MPC is turning off the supplies.
3.
Press
ENTER
to display the first instruction, which reminds you to remove the HV
cables.
4.
Press
ENTER
to display next instruction.
CAL YES
1
2
3
4
5
6
7
8
Set Point
Set Point
Supply 1
Supply 2
Display
Select
Display
Select
Options:
- CAL YES
- CAL NO
5.
Press the top
Display Select
key to calibrate the MPC. If you don't wish to calibrate
at this time, press the bottom
Display Select
key to exit.
RST DFLTS
Digitel MPC Manual
Page 40 of 72