4.
Press the
Menu
key twice to select SUPPLY2 KV, and press
ENTER
. The bottom
display shows the voltage strapped to supply 2.
PRM FACTOR
To display what line voltage the DIGITEL MPC is set to, you need to specify it — 120 or 240.
Use the following procedure.
ATTENTION: Setting the PRM FACTOR merely informs the software how the MPC is
configured, it does not configure it.
1.
Press the
Menu
key to cycle through the menu until you reach CONFIG.
2.
Press
ENTER
to step to the submenu and press the
Menu
key to cycle to PRM
FACTOR.
3.
Press the
ENTER
key.
4.
An arrow points to the bottom Display Select key to show it is active. Press it to
toggle between 120 and 240.
1
2
3
4
5
6
7
8
Set Point
Set Point
Supply 1
Supply 2
Display
Select
Display
Select
arrow shows
bottom key is active
Options:
- 120
- 240
press to toggle
between options
5.
When you reach the desired option, press
ENTER
to save your selection.
Running Diagnostics
Overview
Press the
Menu
key to cycle through the top menu selections until you reach DIAGNOSTCS.
Press
ENTER
to step to its submenu. Press the
Menu
key to cycle through the following 6
submenu selections:
•
REVISION
•
ERROR CODE
•
DISPLY/KEY
•
CALIBRATION
•
RST DFLTS
•
LINE FREQ
Press
ESC
to exit the submenu and return to DIAGNOSTCS.
Digitel MPC Manual