SpectralLED® RS-7 USER MANUAL
Page 61
4.6.7
- LSN - LED board serial number
Reports the serial number as stored in the LED board’s on-board read-only-memory. This serial number cannot
be altered.
Syntax:
LSN
report LED board serial number
Example:
lsn
LHX0152
4.6.8
- USN - RS-7 unit serial number
Reports the serial number as stored in the RS-7 non-volatile memory. This serial number cannot be altered.
Syntax:
USN
report RS-7 unit serial number
Example:
usn
HX2855
4.6.9
– ICK - Integrity Check
Performs an internal check of all RS-7 parameters and calibration data, and verifies that
a)
all monitored internal power supply rail voltages are within tolerance
b)
all parameters are within range & structures are intact
c)
stored calibration data matches the serial number of the LED board for which they were generated
d)
all structures have a valid CRC (CRC32)
If all goes well, only “Ok” is reported. If there are problems, each is reported with a line starting with “ERROR/”
followed by a description. The following is a list of the possible ERROR/ codes:
ERROR/VOLTAGE RAIL: [expected]V measures at [observed]V
ERROR/UNIT STORED PARAMETERS: CRC or structure error
ERROR/UNIT STORED PARAMETERS: appear uninitialized
ERROR/LED BOARD STORED PARAMETERS: CRC or structure error
ERROR/LED BOARD STORED PARAMETERS: appear uninitialized
CH#x ERROR/CALIBRATION DATA: CRC or structure error
CH#x ERROR/CALIBRATION DATA: missing calibration data
CH#x ERROR/CALIBRATION DATA: LED board S/N mismatch