Software Options and Configuration
5-4
COBE
®
Spectra™ Apheresis System •
Essentials Guide
Configuring the AC Infusion Rate
The AC infusion rate is the rate anticoagulant is infused to the donor or
patient. The default AC infusion rate is 0.8 mL/min/L of TBV. You can
configure the AC infusion rate up to 1.1 mL/min/L TBV for any
procedure except TPE.
ELP, LRS, LRST, MNC, PMN, WBC Depletion, and AutoPBSC
Procedures
If you configure the AC infusion rate for any one of these procedures,
the change applies to all these procedures.
For more information and instructions on configuring the AC infusion
rate for platelet collection procedures, see “Configuring the AC Infusion
Rate” in Chapter 2 of the
Platelet Collection Guide
.
For more information and instructions on configuring the AC infusion
rate for cell therapy procedures, refer to the following chapters in the
Cell
Therapy Guide
:
•
MNC procedure – Chapter 2
•
PMN procedure – Chapter 3
•
AutoPBSC procedure – Chapter 4
Configuring Total Plasma Removed (Platelet Collection
Procedures)
You can configure the total volume of plasma removed during a platelet
procedure. This volume can be either a specific volume or a percentage
of the donor’s total blood volume (TBV). For more information and
instructions on configuring this volume, see “Configuring Total Plasma
Removed” in Chapter 2 of the
Platelet Collection Guide
.
Configuring Procedure End Points
ELP, LRS, and LRST Procedures
The default end point for platelet collection procedures is 100 minutes.
For more information and instructions on configuring the end point, see
“Configuring the Platelet Collection Procedure End Point” in Chapter 2
of the
Platelet Collection Guide
.
AutoPBSC Procedures
The default end point for AutoPBSC procedures is 2.0 x TBV. You can
configure this end point based on run time, inlet volume, or number of
TBVs to be processed. For more information and instructions on
configuring the end point, see “Configuring the AutoPBSC End Point”
in Chapter 4 of the
Cell Therapy Guide
.
TPE Procedures
The default end point for TPE procedures is 1.0 plasma volume. For
more information and instructions on configuring the end point, see
“Configuring the TPE Procedure End Point” in Chapter 2 in the
Therapeutic Apheresis Guide
.
MNC Procedures
The default end point for MNC procedures is 2.0 x TBV. You can
configure this end point based on run time, inlet volume, or number of
TBVs to be processed. For more information and instruction on
configuring the end point, see “Configuring the MNC End Point” in
Chapter 2 of the
Cell Therapy Guide
.
Содержание Cobe Spectra
Страница 1: ......
Страница 7: ...COBE Spectra Apheresis System Essentials Guide 1 1 1 Preface ...
Страница 27: ...COBE Spectra Apheresis System Essentials Guide 2 1 2 Introduction ...
Страница 51: ...COBE Spectra Apheresis System Essentials Guide 3 1 3 Installing Transporting and Returning the Spectra System ...
Страница 75: ...COBE Spectra Apheresis System Essentials Guide 5 1 5 Software Options and Configuration ...
Страница 92: ...Software Options and Configuration 5 18 COBE Spectra Apheresis System Essentials Guide ...
Страница 93: ...COBE Spectra Apheresis System Essentials Guide 6 1 6 Automatic and Manual Modes ...
Страница 98: ...Automatic and Manual Modes 6 6 COBE Spectra Apheresis System Essentials Guide ...
Страница 99: ...COBE Spectra Apheresis System Essentials Guide 7 1 7 Alarm Tests ...
Страница 103: ...COBE Spectra Apheresis System Essentials Guide 8 1 8 Anticoagulation ...
Страница 109: ...COBE Spectra Apheresis System Essentials Guide 9 1 9 Fluid Volumes ...
Страница 115: ...COBE Spectra Apheresis System Essentials Guide 10 1 10 General Maintenance ...
Страница 120: ...General Maintenance 10 6 COBE Spectra Apheresis System Essentials Guide ...
Страница 121: ...COBE Spectra Apheresis System Essentials Guide 11 1 11 Helpful Hints and Recovery Procedures ...
Страница 133: ...COBE Spectra Apheresis System Essentials Guide 12 1 12 General Alarms and Troubleshooting ...
Страница 197: ...COBE Spectra Apheresis System Essentials Guide 13 1 13 Specifications ...
Страница 208: ...Specifications 13 12 COBE Spectra Apheresis System Essentials Guide ...
Страница 209: ...COBE Spectra Apheresis System Essentials Guide 14 1 14 Data Input Limits ...
Страница 212: ...Data Input Limits 14 4 COBE Spectra Apheresis System Essentials Guide ...
Страница 213: ...COBE Spectra Apheresis System Essentials Guide 15 1 15 The Seal Safe System ...
Страница 224: ...The Seal Safe System 15 12 COBE Spectra Apheresis System Essentials Guide ...
Страница 225: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 226: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 227: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 228: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 229: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 230: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 231: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 232: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 233: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Страница 234: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...