AK 98 Dialysis Machine - Attentions
B:169
600 Profiling is not restarted
600 Profiling is not restarted
Check UF rate settings and press Confirm.
Appears:
When treatment is resumed after the dialyzer and the blood
lines have been changed during treatment.
Machine actions:
None.
Operator action:
1.
Press the
Fluid
button.
2.
Select the
UF rate
tab.
3.
Press
Profiling
.
4.
Check that the start and stop values for UF rate are the
correct ones and then press
Confirm
. If they are not
correct, deactivate profiling.
Change the values and reactivate profiling. See Section
7
on page A:117 in Operators handbook.
Mute time:
-
Содержание AK 98
Страница 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Страница 45: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 106: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 118: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 128: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 144: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 166: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 196: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 198: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 268: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 275: ...HCEN12745 Revision 10 2016 Program version 2 xx AK 98 Dialysis Machine Attentions B 75 Mute time 2 minutes ...