AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
599 UF rate lower than minimum set
599 UF rate lower than minimum set
Calculated UF rate lower than set min UF rate.
Adjust time and/or UF volume or min UF rate.
Appears:
When the machine needs to decrease the UF rate below
the set minimum UF rate in order to obtain the set UF
volume at the end of treatment.
Machine actions:
Alarm sound is delayed for 2 minutes.
Operator action:
Change one or several of these settings:
●
Decrease the set minimum UF rate
●
Decrease treatment time to increase the UF rate
●
Increase UF volume
Mute time:
2 minutes.
Содержание AK 98
Страница 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Страница 45: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 106: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 118: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 128: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 144: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 166: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 196: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 198: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 268: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 275: ...HCEN12745 Revision 10 2016 Program version 2 xx AK 98 Dialysis Machine Attentions B 75 Mute time 2 minutes ...