B:42
AK 98 Dialysis Machine - Alarms
HCEN12745 Revision 10.2016
Program version 2.xx
223 No BPM values
223 No BPM values
BPM measurement failed. Check BPM cuff and
restart measurement. To clear the alarm press
Confirm.
Appears:
When it was not possible to measure the patient's blood
pressure. The patient may have moved during the
measuring.
Machine actions:
The buzzer sounds.
Mute time: 2 minutes.
Operator action:
1.
To clear the alarm press Confirm.
If the cause of the alarm has not been corrected within
the override time of 2 minutes, the machine alarm
actions will automatically recur.
2.
Make sure that the BPM cuff is correctly wrapped
around the patient’s arm.
3.
Restart measurement.
Содержание AK 98
Страница 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Страница 45: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 106: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 118: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 128: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 144: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 166: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 196: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 198: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 268: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 275: ...HCEN12745 Revision 10 2016 Program version 2 xx AK 98 Dialysis Machine Attentions B 75 Mute time 2 minutes ...