HCEN12745 Revision 10.2016
Program version 2.xx
AK 98 Dialysis Machine - Disinfection with the AK 98 dialysis
machine and WRO system
A:161
2)
Press the
Disinfection
button.
3)
Select the
Heat
tab.
4)
Select the preferred disinfection program.
5)
Press
Confirm
.
6)
If an automatic switch off is required, press
the
On/Off
button for three seconds.
7)
An operator message will be displayed
when the disinfection is completed.
8)
Press the
On/Off
button for three seconds
to switch off the machine.
11.3
Central chemical disinfection program with a WRO unit
11.3.1
Description of central chemical disinfection program with a WRO
unit
The dialysis machine and a WRO water purification unit (WRO 300 H, WRO 300) can
be preset to perform a central chemical disinfection program. Both the dialysis
machine and the WRO unit must be preset by an authorised service technician.
11.3.2
Start a central chemical disinfection with a WRO unit
WARNING!
After a chemical disinfection program, a test for residues shall be performed prior to
connecting a patient to the dialysis machine to make sure that no residues exist in
the dialysis machine.
The following steps are made in the AK 98 dialysis machine.
Procedure
1)
Press the
Disinfection
button.
2)
Select the
Chemical
tab.
3)
Select the preferred central disinfection
program.
The following steps are made in the
WRO unit.
Содержание AK 98
Страница 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Страница 45: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 106: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 118: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 128: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 144: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 166: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 196: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 198: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 268: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 275: ...HCEN12745 Revision 10 2016 Program version 2 xx AK 98 Dialysis Machine Attentions B 75 Mute time 2 minutes ...