A:160
AK 98 Dialysis Machine - Disinfection with the AK 98 dialysis
machine and WRO system
HCEN12745 Revision 10.2016
Program version 2.xx
11.1
General description
The AK 98 dialysis machine and the WRO water purification unit is a system that can
perform integrated heat disinfection and central chemical disinfection. An interface
cable is used to connect the AK 98 dialysis machine with the WRO unit. The cable is
ordered separately.
11.2
Integrated heat disinfection with a WRO 300 H unit
11.2.1
Description of integrated heat disinfection with a WRO 300 H unit
The AK 98 dialysis machine and a WRO water purification unit can be preset to
perform an integrated heat disinfection program. Both the dialysis machine and the
WRO unit must be preset by an authorised service technician.
The integrated heat disinfection procedure includes the following phases:
●
Heat disinfection of the AK 98 dialysis machine.
●
“Low Flow Heat” (LFH), where all wet parts between the AK 98 dialysis machine
and the WRO unit are exposed to hot water at a low flow rate.
●
Heat disinfection of the WRO 300 H unit, if this has been selected as preset on the
WRO unit.
The dialysis machine can be set to start a scheduled integrated heat disinfection.
11.2.2
Schedule an integrated heat disinfection
Procedure
1)
Check that the WRO 300 H unit is
switched on.
2)
Press the
Disinfection
button.
3)
Select the
Heat
tab.
4)
Press
Heat
or
Heat + LFH
.
11.2.3
Start an integrated heat disinfection manually
Procedure
1)
Check that the WRO 300 H unit is
switched on.
The following steps are made in the
AK 98 dialysis machine.
Содержание AK 98
Страница 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Страница 45: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 106: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 118: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 128: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 144: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 166: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 196: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 198: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 268: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 275: ...HCEN12745 Revision 10 2016 Program version 2 xx AK 98 Dialysis Machine Attentions B 75 Mute time 2 minutes ...