A:146
AK 98 Dialysis Machine - Disinfection and cleaning
HCEN12745 Revision 10.2016
Program version 2.xx
NOTE!
For the best cleaning results, start with a decalcification.
10.3.3
Start a heat disinfection
Procedure
1)
Press the
Disinfection
button.
2)
Select the
Heat
tab.
3)
Press
Heat
.
4)
Press
Confirm
.
10.3.4
Start a heat disinfection with a CLEANCART cartridge
Procedure
1)
Press the
Disinfection
button.
2)
Select the
Heat
tab.
3)
Press
Heat CleanCart
.
4)
Press
Confirm
.
5)
When you are prompted by an attention:
Attach the CleanCart cartridge to the
holder.
10.3.5
Start a heat disinfection with liquid citric acid
Procedure
1)
Press the
Disinfection
button.
2)
Select the
Heat
tab.
3)
Press
Citric 20%
.
4)
Make sure that the disinfectant inlet tube is
placed in the disinfectant solution.
Содержание AK 98
Страница 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Страница 45: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 106: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 118: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 128: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 144: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 166: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 196: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 198: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 268: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 275: ...HCEN12745 Revision 10 2016 Program version 2 xx AK 98 Dialysis Machine Attentions B 75 Mute time 2 minutes ...