AK 98 Dialysis Machine - Measuring blood pressure
A:131
Results
The latest measured value is available in the BPM read out field.
The dialysis machine measures blood pressure at the set intervals, starting when the
first interval has passed. For example, if you set to measure with 20 minutes intervals,
the first measuring will start after 20 minutes.
It is possible to start a single measurement at any time between the intervals if
necessary. The next measurement will then be delayed and start 20 min after the
single measurement (if 20 min intervals is set).
When a single measurement is running or auto mode is activated no BPM alarms will
be generated if the BPM alarm off symbol is lit in the status bar.
To stop an ongoing measurement: press the
BPM
button to open
Blood Pressure
Menu
and press
Blood pressure measurement
or press the readout field on main
screen.
When treatment has ended, and the cuff has been removed from the patient:
●
The blood pressure measurement will continue as set, when automatic blood
pressure measuring has been activated.
●
When the next functional check is started, the machine will return to displaying
default values.
8.5
Measurement history
The blood pressure measurments are logged and saved during the treatment. The
measurement data is viewed in the
History
tab.
Procedure
1)
Press the
BPM
button.
2)
Select the
History
tab.
3)
Press
History
.
8.6
Set alarm limits
It is possible to set high and low alarm limits for systolic, diastolic and mean blood
pressure as well as pulse rate.
Содержание AK 98
Страница 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Страница 45: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 106: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 118: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 128: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 144: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 166: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 196: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 198: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 268: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 275: ...HCEN12745 Revision 10 2016 Program version 2 xx AK 98 Dialysis Machine Attentions B 75 Mute time 2 minutes ...