HCEN12745 Revision 10.2016
Program version 2.xx
AK 98 Dialysis Machine - Profiling
A:125
7.6
Set and activate profiling with a preset model
Procedure
1)
Press the
Fluid
button.
2)
Select the
UF rate
tab.
3)
Press
Profiling
.
4)
Press a preset model,
Model 1
,
Model 2
or
Model 3
.
5)
Check that the preset values in the model
are correct.
6)
Activate the profiling model. If desired, it is
possible to activate the profiling function at
any time during priming or treatment.
The parameter settings of the complete
model will be activated simultaneously. It is
possible to adjust the preset values of the
parameters when the profiling model is
deactivated, refer to Section 7.5.1
Содержание AK 98
Страница 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Страница 45: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 106: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 118: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 128: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 144: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 166: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 196: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 198: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 268: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 275: ...HCEN12745 Revision 10 2016 Program version 2 xx AK 98 Dialysis Machine Attentions B 75 Mute time 2 minutes ...