A:115
6.2.2
How to add a second and subsequent phase of isolated UF
Second and subsequent phases of isolated UF can be activated at any point during
the treatment.
The time and UF volume for the new phase must be added to the previous isolated UF
time and UF volume to create a cumulative amount.
Example: If the 1st phase was set to 30 minutes and 0.5L and a second phase is to be
set with the same values, the settings need to be 60 minutes and 1.0L which is the
total of the two phases together.
The values set for the second and subsequent phases must always be larger than the
accumulated isolated time and UF volume.
6.2.3
Deactivate isolated ultrafiltration
If necessary, isolated UF can be deactivated at any time during the phase.
Procedure
1)
Press the
Fluid
button.
2)
Press
Isolated UF
.
3)
Deactivate Isolated UF.
6.3
Additional information
6.3.1
Heparin
If the heparin pump is used, it will run during the complete isolated UF, unless isolated
UF is performed after the diffusion phase. Settings for
Heparin bolus volume
,
Heparin flow rate
and
Stop time
are kept.
Содержание AK 98
Страница 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Страница 45: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 106: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 118: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 128: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 144: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 166: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 196: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 198: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 208: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 268: ...This page is intentionally left blank This page is intentionally left blank ...
Страница 275: ...HCEN12745 Revision 10 2016 Program version 2 xx AK 98 Dialysis Machine Attentions B 75 Mute time 2 minutes ...