Galvanic Applied Sciences USA, Inc.
41 Wellman St.
Lowell, MA. 01851
Tel: 978-848-2701
Fax: 978-848-2713
Email: [email protected]
45
Technical data model MoniTurb-S (90° scattered light)
(Illustration similar)
•
Principle of measurement: 90° side scattered light
•
Calibration interval 24 months
•
Colour compensation
•
Cleaning in place
•
Flow cell material: 316SS/ 1.4404 (other materials on request)
•
Protection: NEMA 4X / IP65
•
Process connections: DIN, ANSI, SMS, NPT, APV, TH, ...
•
Line sizes DN 25 to DN 125 (1/2” to 4“)
•
Process pressure: PN16 / ANSI class 150
•
Process temperature: max. 140° C
•
Cable length: max. 100 m (approx. 300 feet)
•
Optional cleaning jets
•
Optional hazardous area Ex-zone I and Ex-zone II
•
Compatible to converter type: Messenger
Applications
The Monitek sensor Model MoniTurb-S, in connection with a Monitek converter type Messenger,
measures turbidity in liquids. The system can be used for inspection, control and optimisation of industrial
processes:
Drinking water
Waste water
Chemical industry
Paper industry
Biotechnology
Beverages
Ultra filtrate
Filtrate
Unfiltrate
Condensate
Process water
Water in oil / oil in water
Description
The turbidity sensor Model MoniTurb-S uses the principle of 90° side scattered light to detect
suspended particles in liquids. The ratio of scattered light/direct light assures highly reliable
and repeatable measurement results. The sensor is specially designed to detect small particle
sizes (high sensitivity for particles < 0,3
µ
m). The sensor can be installed
into almost any type of pipe. Process connection, pressure, temperature, gasket material, etc
will be application specific.
The optional feature for hazardous area, zone I or zone II, allows for a wide range of
applications. A maximum cable length of 100 m (approx. 300 feet) allows the converter to be
located, outside the hazardous area.
Specifications
Line sizes:
DN 25 to DN 125 mm / 1/2“ to 4“
Optical path length:
Application specific
Process pressure:
PN 16 / ANSI class 150
Measurement range:
0 – 500 ppm
Process temperature: Max. 140° C
Detector system:
Silicone diodes
Sensor material:
316SS / 1.4404 (other materials on request)
Hazardous area (optional): Ex-zone I and Ex-zone II
Gasket material:
Application specific
Cleaning jets: