Maintenance
Replacement of measurement lamp
In
tro
du
ct
io
n
Ignoring of the following hints will cause a loss of warranty
Qualified technical personnel must perform repair and maintenance.
Before beginning any work the sensor must be cleaned and flushed thoroughly. Depending
on customers’ application product residues can be very dangerous (aggressive, poisonous).
Please handle the system very careful due to possible leakage etc.
Avoid pulling and twisting of the lamp cable.
Avoid applying excessive force during assembling and disassembling of the sensor.
Screw in all bolts and cable glands stalwart only.
Please work carefully during the replacement of the measurement lamp.
The lamp replacement should be done in a dry and clean location, to protect the optical
components from dirt.
Please be sure that no dust or other particles penetrates into the optical assembly.
Do not touch the lenses.
In case of dirty component please clean carefully by using fresh water and a cloth (lint-free),
dry all components by using instrument air.
Use suitable tools only.
Required tools:
-
Special Monitek lamp spanner wrench
-
2 x open end wrench - 22 mm
-
Flat blade screwdriver - 2 mm
-
Phillips screwdriver – medium size
-
Hexagonal spanner – 1.5 mm
-
Hexagonal spanner – 2.5 mm
-
Needle nose pliers – small
-
Small bowl to drop components
Pic
tu
re
1
The lamp cable is marked with “lamp”
1.
Please loose the upper hexagonal nut of the cable gland
by using a 22 mm open-end wrench. Use the second
22 mm open-end wrench to hold the lower hexagonal
nut of the cable gland in position.
This will assure that you do not twist the lamp cable.