ViscoSite VL800 Viscometer
Operating Manual
ViscoSite VL800 Viscometer
Galvanic Applied Sciences, Inc.
Email: [email protected]
Phone: (978) 848--2701
Operating Manual Rev: 2.0E
Toll-Free: 1 (866) 252-8470
Page 67 of 132
The relay outputs may also be configured using the
ViscoSite software. See Section .
4.6 Option Setup Menu
Accessing the Option Setup Submenu
•
Press the Option Setup button on the keypad;
•
The Option submenu shown in Figure 4.10 is displayed.
The Option submenu allows for the configuration of critical parameters
associated with analog output, behavior, relay alarm behavior, and
data processing performed by the ViscoSite system.
Figure 4.10: Option Submenu (Partial)
Density Mode, Nominal Process Temperature
(Temp Comp), Time Averaging Mode, Averaging
Time
The list of parameters in the Option submenu, their configurable
settings, and their functional descriptions are shown in Table 4.8.
Parameter Options
Explanation
Density
Auto Density
Enabled
The ViscoSite system is configured to
receive density information via a real-
time 0-20mA / 4-20mA isolated analog
input signal from an external density
transducer.
Manual
Density
Enabled
A single representative density value
is manually entered into and stored by
the ViscoSite transmitter (See Section
5.8.1).
Nom
Temp
Nominal
Temperature:
User Entered
Numerical
The temperature of the process at
normal operating conditions. This
value, along with other information, is
used for calculating temperature