![Galvanic Applied Sciences ViscoSite VL800 Скачать руководство пользователя страница 32](http://html1.mh-extra.com/html/galvanic-applied-sciences/viscosite-vl800/viscosite-vl800_operating-manual_3300670032.webp)
ViscoSite VL800 Viscometer
Operating Manual
ViscoSite VL800 Viscometer
Galvanic Applied Sciences, Inc.
Email: [email protected]
Phone: (978) 848--2701
Operating Manual Rev: 2.0E
Toll-Free: 1 (866) 252-8470
Page 31 of 132
Installation Considerations for the Transmitter
The ViscoSite transmitter is designed for operation at temperatures
from 0° C to 60° C.
The transmitter should be mounted in a location where it is
•
not exposed to direct sunlight or rainfall
•
not exposed to excessive vibration
•
within 300 metres of cable run to the transducer, taking into account
indirect pathways
Installation Considerations and Space
Requirements for the Transmitter Enclosure
The dimensions of the ViscoSite transmitter enclosure are shown in
Figure 3.6.
Figure 3.6: ViscoSite Transmitter Enclosure with Dimensions
Be sure to leave space
below
the transmitter enclosure to allow for the
connection of signal, power, and communication wires.
Be sure to leave space to the
left
of the transmitter enclosure to allow
the enclosure door and swing panel to fully open.