2.5.2 Pulse Counter
The pulse counter is used to ensure even spacing of stains on the sensing tape. The pulse
counter is shown in Figure 2-4.
Figure 2-4: Pulse Counter
The left and centre pictures show the pulse counter as it appears when not installed in the
chassis and the right picture shows how the pulse counter appears in the tape transport
system. The black aluminum (or grey plastic) capstan is mounted on the silver nub seen in
the left picture, and it rotates when the tape moves over it.
The pulse counter produces a specific amount of pulses for every revolution. By measuring
the number of pulses produced by the movement of the tape while the motor is running, the
system processor can ensure that the tape is advanced exactly the same distance before
every analysis cycle, regardless of the quantity of tape on the take-up reel. This ensures that
the stain spacing on the tape is even; further, by reducing the amount of white space
between subsequent stains, tape wastage is minimized, and tape life is maximized.
When the motor advances the take-up reel, causing the tape to move, the movement of the
tape causes the capstan on the pulse-counter to rotate, producing pulses. These pulses are
read by the sensor block, which then sends the pulse count to the system processor. When
the processor receives a number of pulses equal to the number of pulses per advance set in
the application program; typically around 40 pulses, or a little less than one sixth of a
revolution, it sends a signal to the motor to stop turning, and the tape stops advancing.
The pulse counter also allows the system to accurately indicate the quantity of tape
remaining on the supply reel, as long as the initial length is set correctly in the application
program or through the keypad. This value is displayed on the LCD in the upper right corner,
and can be output through an analog output. The quantity of tape remaining is calculated
after every analysis cycle, using the formula outlined in Section 5.9.1.1. This value can also
be used to trigger a low-tape alarm when the quantity of tape goes below a user-set value,
set up through the application program.
Содержание ProTech903W
Страница 8: ...Revision 1 8 9 10 2019...
Страница 109: ...Revision 1 109 9 10 2019 Figure 5 34 Create Modbus List Dialogue Box...
Страница 112: ...Revision 1 112 9 10 2019 Figure 5 36 Typical Modcon with Floating Point List...
Страница 141: ...Revision 1 141 9 10 2019 Figure 6 18 Status Tab...
Страница 158: ...Revision 1 158 9 10 2019 Figure 10 1 DC Power Wiring Diagram Figure 1010 2 AC Power Wiring Diagram...
Страница 159: ...Revision 1 159 9 10 2019 Figure 10 3 External Furnace Wiring Diagram Figure 10 4 AC DC Power Wiring Diagram...
Страница 160: ...Revision 1 160 9 10 2019 Figure 10 5 Solenoid Drivers Wiring Diagram Figure 10 6 Relays Wiring Diagram...
Страница 164: ...Revision 1 164 9 10 2019 Figure 10 13 Isolated RS 485 Port Figure 10 14 P17 RS232 Port to Ethernet Cable...
Страница 171: ...Revision 1 171 9 10 2019...
Страница 172: ...Revision 1 172 9 10 2019 Section 13 MSDS Information...