3.1 Receiving the System
When the system arrives, inspect the packaging for external signs of damage. If there is any
obvious physical damage, contact the shipping agent and Galvanic Applied Sciences to
report the damage and request that the carrier's agent be present when the unit is unpacked.
It is recommended that you retain the shipping container so that it may be used for future
shipment of the unit, if necessary.
3.2 Installation Requirements
3.2.1 Electrical Requirements
The system can be powered by 100-230 VAC or 10-32 VDC sources. The power requirement
is 10 W.
If desired, it is possible to connect both AC and DC power supplies. In this manner, the DC
power supply as a backup power. When choosing a power source, the type of solenoid
drivers need to be considered (AC or DC).
The total sulfur option requires 250 W at 120 VAC or 230 VAC.
3.2.2 Location of the System
The system is designed to be operated at ambient temperatures from 10 - 50
o
C. Galvanic
offers complete analyzer shelters from sun shades to complete building, please contact
Galvanic Applied Sciences, Inc. (or your local representative) for additional information.
The system should be mounted in a location where it is not exposed to excessive vibration.
In addition, it should be mounted in a location where the ambient pressure remains relatively
consistent over the course of a day
The vent line cannot have any backpressure on it, as backpressure can seriously impact
readings. The vent line should be as short as possible at a slight continuous downward slope.
3.2.3 Space Requirements
The system dimensions are provided in Figure 3-1 for Class I systems and Figure 3-2 for
Class I Division 1 systems with the total Sulfur option. For Class I Division 2 systems, the
corresponding figures are presented in Figures 3-3 and 3-4
Содержание ProTech903
Страница 2: ......
Страница 96: ...Revision 14 96 August 02 2019 Figure 5 19 Event Log...
Страница 122: ...Revision 14 122 August 02 2019 Figure 5 38 Typical Modicon with Floating Point List...
Страница 124: ...Revision 14 124 August 02 2019 Figure 5 40 Expanded Modbus Nodes...
Страница 167: ...Revision 14 167 August 02 2019 Figure 9 1 DC Power Wiring Diagram Figure 9 2 AC Power Wiring Diagram...
Страница 168: ...Revision 14 168 August 02 2019 Figure 9 3 Total Sulfur Wiring Diagram Figure 9 4 AC DC Power Wiring Diagram...
Страница 169: ...Revision 14 169 August 02 2019 Figure 9 5 Solenoid Drivers Wiring Diagram Figure 9 6 Relays Wiring Diagram...
Страница 173: ...Revision 14 173 August 02 2019 Figure 9 13 Isolated RS 485 Port Figure 9 14 P17 RS232 Port to Ethernet Cable...
Страница 175: ...Revision 14 175 August 02 2019 Figure 9 17 Non Isolated 4 20 mA Inputs...
Страница 190: ...Revision 14 190 August 02 2019...