-30-
actually re-plot the data and zoom in or out on the y-axis by entering different
values.
2.5 Self-Cleaning Interval
The interval between automatic self-cleaning cycles is selected from the “Set
Cleaning Event” option on the System menu. The value is entered in minutes.
The MONITEK MESSENGER holds the display and all outputs at the previous
value while the self-cleaning cycle is underway. Once the self-cleaning cycle is
complete, the instrument returns to normal operation. The self-cleaning cycle
can always be initiated or tested by selecting the “Start Clean Cycle” option from
the Set Cleaning Event Menu.
When manually starting a cleaning cycle make sure you have the variables set
properly and a device selected.
Note:
The Monitek Messenger Provides the electronic control for the signals
needed to activate cleaning jets. The user must provide the appropriate
hardware, including sensor with cleaning jets, a solenoid valve to switch either
pressurized air or cleaning solution (water for example) to the cleaning jets in the
flow cell.