Revision 5
57
August 2021
Take note of the following rules-of-thumb when comparing the pink trace of the factory
reference with the red trace of the current transmission spectrum:
•
The factory reference should be largely similar to the current transmission spectrum
when the analyzer is in back purge mode. If it is not, there is likely an issue with the
analyzer optical system. Refer to Section 7.2 for troubleshooting techniques.
•
The factory reference should be notably different from the current transmission
spectrum, as it is in Figure 34, when the analyzer is in sampling mode. If the sampling
spectrum is the same or very similar to the factory reference, this could indicate an
issue with plugging in the sample probe. Refer to Section 7.9 for details on how to clear
a plugged probe.
The
New Reference
button is used to make the analyzer use the currently displayed
transmission spectrum to calculate a new zero baseline without carrying out a full zero
calibration cycle. Highlight the New Reference button by using the FIELD NEXT / FIELD
PREV keys, then pressing ENTER.
Pressing
New Reference
DOES NOT cause the analyzer to automatically go into
back purge mode. Be sure the analyzer is in back purge mode prior to using this
function otherwise the calculated analyzer results after calculating the new
reference WILL NOT BE VALID!
4.7
Absorbance Panel
The
Absorbance
panel shown in Figure 35 shows the currently calculated absorbance
spectrum as well as the current concentration of all of the analyzed components in the tail gas
stream.
Figure 35: Absorbance Panel
Содержание 943-TGXeNA
Страница 130: ...Revision 5 130 August 2021 Figure 82 Oven Cabinet Door Removed ...
Страница 131: ...Revision 5 131 August 2021 Figure 83 Control Cabinet Door Removed ...
Страница 132: ...Revision 5 132 August 2021 Figure 84 Power Steam Air Signals Connection Details ...
Страница 133: ...Revision 5 133 August 2021 Figure 85 AC Wiring Schematic ...
Страница 134: ...Revision 5 134 August 2021 Figure 86 DC Signals and Wiring Diagram ...
Страница 135: ...Revision 5 135 August 2021 Figure 87 Flow Diagram ...