concentration outputs are configured for Track, though this is entirely up to the user to
decide. The value output by each analog output is updated once per second. The
configuration of the analog outputs can be done either using the local display and keypad,
as described in Section 4.8.1, or by using the Web GUI, as described in Section 5.4.
3.3.2 Digital Outputs (Relays)
There are four digital outputs
on the analyzer which are used for outputting analyzer status
and other operational condition data. These digital outputs are all zero potential form C relay
contacts, each isolated to 250 V
rms
. The four status and operational relays are
Status
,
Service
,
Mode
, and
Control
. Each of these relays has an associated status indicator on the
analyzer display for visual confirmation of the analyzer status. In addition to the 4 digital
outputs associated with the analyzer operation, there is an independent
Loss of Purge
signal output connected to the purge control box on top of the analyzer control cabinet. The
function of each relay is described in the following sections.
3.3.2.1
Status
The
Status
output indicates whether there are any faults present that will force the
analyzer into back-purge mode until the fault is cleared. Table 3 describes the Status
relay states.
Table 3: Status Relay States
State
Display Indicator Colour
Explanation
Normal
Green
All monitored system parameters are
within normal limits. No faults are present
Fault
Red
At least ONE of the following conditions
is true:
•
Analyzer power is off
•
Measurement Cell temperature
is outside specified range
•
Probe temperature is outside
specified range
NOTE: Only for
analyzers equipped with
probe temperature RTD
•
Measured sulfur vapour
concentration is above specified
fault setpoint
•
Spectrometer parameters are
outside specified range
•
The system computer is unable
to communicate with the I/O
board
•
The system computer is unable
to communicate with the
spectrometer
An analyzer with at least one fault condition present will be unable to carry out sampling
and analysis of the process gas. When the fault condition clears, the analyzer will
automatically perform an auto-zero calibration and then return to normal sampling mode.
Содержание 943-TGXeNA
Страница 130: ...Revision 5 130 August 2021 Figure 82 Oven Cabinet Door Removed ...
Страница 131: ...Revision 5 131 August 2021 Figure 83 Control Cabinet Door Removed ...
Страница 132: ...Revision 5 132 August 2021 Figure 84 Power Steam Air Signals Connection Details ...
Страница 133: ...Revision 5 133 August 2021 Figure 85 AC Wiring Schematic ...
Страница 134: ...Revision 5 134 August 2021 Figure 86 DC Signals and Wiring Diagram ...
Страница 135: ...Revision 5 135 August 2021 Figure 87 Flow Diagram ...